1.
Chinese Journal of Medical Instrumentation ; (6): 371-373, 2020.
Article in Chinese | WPRIM | ID: wpr-828184

ABSTRACT

This paper introduces a fast repair methods of Versa HD volume-modulated accelerator's high voltage circuit fault:the key points test method. To identify five key points:①Enter maintenance mode to check for AFC deviation; ② The magnetron waveform MI and PFN waveform PFN V are detected on the maintenance terminal board; ③ Detect thyratron waveform; ④ Check the supply voltage of thyratron; ⑤ HT PSU 600 V DC test 600 V normal or not. The waveform or voltage is also measured to efficiently narrow the fault range to find out the cause of the fault, quickly remove the fault.


Subject(s)
Maintenance
SELECTION OF CITATIONS